How do I use the CO2 scrubber for calibrations?

  1. Remove the end caps from each side of the CO2 Scrubber.

  2. Go to the calibration screen of the PGA and turn the pump ON.

  3. Attach either end of the CO2 scrubber to the inlet tube of the PGA.

  4. Watch the zero calibration values until they drop to near 0.00%.

  5. Select “Start Calibration” and allow the green progress bar to start.

  6. Once complete, turn the pump off and then connect to your Span calibration gas.