SuperSystems

SSI_trans_small3

Continuous NDIR-MGA Single Furnace

NDIR MGA Single Furnace Analyzer: SuperSystems Precision

 Permanently mounted and continuous Precision Measurement of Protective Heat Treating Atmospheres

CO, CO2, CH4, H2 – Used for Carburizing, FNC, and Nitriding

The permanently-mounted continuous gas analyzer can be used for carburizing, neutral and ferritic nitrocarburizing heat treat applications. The measurement of the process gas provides calculated values for %C, KN, KC1.  In endothermic based atmospheres, the instrument in conjunction with carbon probes can be used to automatically adjust COF to provide the most precise atmosphere measurement.

1 – Calculation of Kc may have inaccuracies in the presence of ammonia.

Enhances quality through process repeatability and accuracy
On-screen trending chart
Auto adjust mode replaces manual adjustment of COF – Probe+IR
Sooting, air leaks, water leaks, and radiant tube leaks quickly identified through 3 gas measurement
Auto-Calibration integrated into the MGA 6010
Session recording with customizable user names and equipment
FNC and Nitriding Calculations with optional H2 Cell
 
Sampling methodExtraction by internal pump (when necessary)
Measurement Method (CO, CO2, CH4)Non-Dispersive Infrared (NDIR)
Measurement Method (H2)Thermal Conductivity
Accuracy and repeatability± 1% of full scale
Recommended Factory Calibration IntervalAnnual
Flow MeterFront of panel and also on-screen
AC Power Requirements90 to 230 VAC, 50 to 60 Hz, 60 Watts
CommunicationsEthernet, USB(A), USB(B), RS485 Modbus
Data StorageContinuous automatic data logging
Data RetrievalXGA Viewer Software (included) or on-screen
Operating Temperature32° to 122° F (0° to 50° C)
CO: Carbon MonoxideRange: 0 – 30%
CO2: Carbon DioxideRange: 0 – 2.0%
Optional Range: 0 – 20.0%
CH4: Natural Gas/MethaneRange: 0 – 20%
Optional H2: HydrogenRange: 0 – 100%
Calculated % CarbonRange: 0.1 – 2.00%
Suggested COF / PF factors 
Onboard Data Acquisition 
Enhanced with Nitriding/FNC CalculationsCarbon Potential (KC)
Nitriding Potential (KN)
COF limits predetermined for % Carbon adjustment
Millivolt and temperature limits for furnace gas sampling
Switch from probe + IR control to “monitor” mode